1290 Infinity Quaternary Pump User Manual 147
Maintenance
9
Installing Fittings and Capillaries
Installing Fittings and Capillaries
1 Install fittings and capillaries.
2 Tighten fittings and capillaries.
WARNING
Solvent can spray under high pressure.
Observe appropriate safety procedures (for example, goggles, safety gloves and
protective clothing), when opening flow path.
CAUTION
Deformation of fittings and seals
Liquid drops under high pressure act like solid parts. Tightening connections under
high pressure can deform or destroy fittings and seals.
Never tighten flow connections under pressure.
NOTE
The lifetime of a fitting depends on how firmly it has been tightened; firm tightening
reduces the lifetime.
If fitting has been overtightened, replace it.