Error Information
Pump Error Messages
Drive phases differ too much in electric resistance
Error ID: 29216
Pump drive calibration has failed due to a strong difference electric
resistances for different motor phases.
Degasser's pressure limit violation
Error ID: 29220
Pressure too far above the limit.
Probable cause Suggested actions
1
Pump drive cable defect.
Please contact your Agilent service
representative.
2
Pump drive defect.
Please contact your Agilent service
representative.
3
Defective main board.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1
Leak in degasser chamber or degasser
tubing.
Please contact your Agilent service
representative.
2
Defect vacuum pump.
Please contact your Agilent service
representative.
3
Degasser chamber empty or connected to
air.
Block unused degasser channels.

Table of Contents

Other manuals for Agilent Technologies 1290 infinity

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1290 infinity and is the answer not in the manual?

Agilent Technologies 1290 infinity Specifications

General IconGeneral
BrandAgilent Technologies
Model1290 infinity
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals