130 1290 Infinity Quaternary Pump User Manual
7
Error Information
Pump Error Messages
Drive phases differ too much in electric resistance
Error ID: 29216
Pump drive calibration has failed due to a strong difference electric
resistances for different motor phases.
Degasser's pressure limit violation
Error ID: 29220
Pressure too far above the limit.
Probable cause Suggested actions
1
Pump drive cable defect.
Please contact your Agilent service
representative.
2
Pump drive defect.
Please contact your Agilent service
representative.
3
Defective main board.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1
Leak in degasser chamber or degasser
tubing.
Please contact your Agilent service
representative.
2
Defect vacuum pump.
Please contact your Agilent service
representative.
3
Degasser chamber empty or connected to
air.
Block unused degasser channels.