1290 Infinity Quaternary Pump User Manual 113
Error Information
7
General Error Messages
Compensation Sensor Short
Error ID: 0080
The ambient-compensation sensor (NTC) on the power switch board in the
module has failed (open circuit).
The resistance across the temperature compensation sensor (NTC) on the
power switch board is dependent on ambient temperature. The change in
resistance is used by the leak circuit to compensate for ambient temperature
changes. If the resistance across the sensor falls below the lower limit, the
error message is generated.
Probable cause Suggested actions
1
Defective power switch assembly
Please contact your Agilent service
representative.
2
Loose connection between the power
switch board and the main board
Please contact your Agilent service
representative.
3
Defective main board.
Please contact your Agilent service
representative.