152 1290 Infinity Quaternary Pump User Manual
9
Maintenance
Replacing the Inlet Weaver
3
Pull the inlet weaver out of the inlet valve.
4
Insert the new inlet weaver to the inlet valve. Fix the
weaver with the plastic screw.
5
Fix the fitting of the new inlet weaver to the MCGV.
6
Reconnect tubings between MCGV and degasser.
)LWWLQJ