158 1290 Infinity Quaternary Pump User Manual
9
Maintenance
Replacing the Outlet Valve
5
Insert the heat exchanger capillary into the outlet of the
outlet valve. Using a torque wrench with a 2.5 mm hex
bit, set 3 Nm and close the hex screw at the top of the
outlet.
6
Purge and condition the system to remove air.
1P