38 1290 Infinity Quaternary Pump User Manual
3
Installing the Module
Installation Information on Leak and Waste Handling
1 Stack the modules according to the adequate stack configuration.
The leak pan outlet of the upper module must be vertically positioned above
the leak tray of the lower module, see Figure 11 on page 37.
2 Connect data and power cables to the modules, see section Installing the
Module below.
3 Connect capillaries and tubes to the modules, see section Flow Connections
to the module below or the relevant system manual.
WARNING
Toxic, flammable and hazardous solvents, samples and reagents
Keep solvent path free from blockages.
Keep the flow path closed (in case the pump in the system is equipped with a
passive inlet valve, solvent may leak out due to hydrostatic pressure, even if your
instrument is off).
Avoid loops.
Tubes must not sag.
Do not bend tubes.
Do not immerse tube end in waste liquid.
Do not intubate tubes in other tubes.
For correct tubing follow instructions on label attached to the module.

Table of Contents

Other manuals for Agilent Technologies 1290 infinity

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1290 infinity and is the answer not in the manual?

Agilent Technologies 1290 infinity Specifications

General IconGeneral
BrandAgilent Technologies
Model1290 infinity
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals