72 1290 Infinity Quaternary Pump User Manual
4
Using the Module
Setting up the Pump with the Instrument Control Interface
Blend Assist
The Blend Assist table allows you to blend two or more solvents or solvent
mixtures from stock solutions. The blends must be of pure solvents or pure
solvents with additives. For example, you can blend 100 % water with 10 %
isopropanol in water.
Channel: The channel name.
•Type: The type of solvent
Solvent <n>: Pure solvent
Solvent <n> Additive: Solvent mixture
Calibration: Click the down arrow and select the solvent or solvent mixture
from the list.
•Name: Enter a name for the solvent or solvent mixture in this field.
Stock conc.: For solvent mixtures, specify the concentration of the additive
in the stock solution in this field. Pure solvents are always 100 %.
Final conc.: Enter the concentration of the additive that you want to achieve
in this field. The pure solvent and solvent mixture will be blended to
achieve the Final conc. For the relationship of stock concentration and
concentration in the mixture, the composition accuracy needs to be
considered (see “Performance Specifications” on page 26).
Conc. unit: The concentration can be defined as mM (mmol/L) or as %.

Table of Contents

Other manuals for Agilent Technologies 1290 infinity

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1290 infinity and is the answer not in the manual?

Agilent Technologies 1290 infinity Specifications

General IconGeneral
BrandAgilent Technologies
Model1290 infinity
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals