EasyManua.ls Logo

Agilent Technologies 1290 infinity - Page 98

Agilent Technologies 1290 infinity
300 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
98 1290 Infinity Quaternary Pump User Manual
5
How to Optimize the Performance of Your Module
How to Achieve Higher Resolution
This shows that k and hence resolution can be increased by having a shallower
gradient (2 to 5 %/min change is a guideline), higher flow rate and a smaller
volume column. This equation also shows how to speed up an existing gradient
– if the flow is doubled but the gradient time is halved, k
*
remains constant
and the separation looks the same but happens in half the time. Recently
published research has shown how a shorter STM column (at temperatures
above 40 °C) can generate higher peak capacity than a longer STM column by
virtue of running it faster. (Refer to Petersson et al., J.Sep.Sci, 31, 2346-2357,
2008, Maximizing peak capacity and separation speed in liquid
chromatography).

Table of Contents

Other manuals for Agilent Technologies 1290 infinity

Related product manuals