Automatic Liquid Sampler and Injector
Syringes
Released: March 2004 6850 Series II GC User Information page 225 of 256
Figure 44. Removing the needle support assembly
4. With your finger under the shaft near the bearing on the needle support
assembly, pull gently to release the bearing from the bearing clip in the
syringe carriage.
5. Carefully use the bearing to pull the rod down until you can lift the
assembly out of the syringe carriage.
Caution Be careful not to pull the assembly by its metal shaft. The shaft is easily bent.
Bearing clip
Guide
Syringe
Slide
Bearing
Syringe carriage
Plunger carrier loop
carriage
tracks