121
Repairing the Autosampler
Metering Seal and Plunger
2 Remove the two fixing bolts, and pull the
head assembly away from the sampler.
3 Remove the two fixing bolts from the base of
the metering head assembly.
4 Disassemble the metering head assembly.
5 Use a small screwdriver to carefully remove
the seal. Clean the chamber with lint-free cloth.
Ensure all particular matter is removed.
6 Install the new seal. Press the seal firmly into
position.
7 Place the piston guide on top of the seal.