EasyManua.ls Logo

Agilent Technologies G1313A - Injection Sequence

Agilent Technologies G1313A
292 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
205
Introduction to the Autosampler
Sampling Sequence
Injection Sequence
Before the start of the injection sequence, and during an analysis, the
injection valve is in the mainpass position (Figure 34). In this position, the
mobile phase flows through the autosamplers metering device, sample
loop, and needle, ensuring all parts in contact with sample are flushed
during the run, thus minimizing carry-over
Figure 31 Mainpass Position
When the sample sequence begins, the valve unit switches to the bypass
position (Figure 32). Solvent from the pump enters the valve unit at port 1,
and flows directly to the column through port 6.
Figure 32 Bypass Position

Table of Contents

Related product manuals