Troubleshooting
i3070 Series 5i Help 6-7
The drawer is not in the lowered position.
Check the drawer’s Left Drawer Down Sensor (PHS-23) and/or
Right Drawer Down sensor (PHS-21).
Lower the drawer.
Disengage the front safety hook.
Check the front hook sensor (PHS-17). See Sensor Troubleshooting.
Disengage the rear safety hook.
Check the rear hook sensor (PHS-16). See Sensor Troubleshooting.
A board was removed when the system is running in Auto mode.
1 Restart the testplan.
A board is at the exit position of Zone 2 but no status is received from
the controller.
Remove the board and restart the testplan.
Table 6- 2 Error Messages
Error Message Cause/Action