Spectrum Analyzer Commands 4-2 :CALCulate Subsystem
S412E PM PN: 10580-00319 Rev. N 4-11
:CALCulate:LIMit:VALue <value>
Title: Move Limit
Description: Sets the currently active limit line value. This command moves an
entire single or multi-segment limit line up or down by the given
<value>. This command is equivalent to the command
:CALCulate:LIMit:Y. Use :CALCulate:LIMit:TYPe to set the currently
active limit line.
Note that this will change the Move Limit on the front panel to Limit if
it is currently set to Point.
Parameter: <value>
Default Unit: Current y-axis unit.
Related Command: :CALCulate:LIMit:Y
:CALCulate:LIMit:TYPe
Front Panel Access: Shift-6 (Limit), Limit Edit, Amplitude
:CALCulate:LIMit:VERTical:ADD
Title: Add Vertical
Description: Adds vertical. This will add two inflection points that share the same
frequency and are centered midpoint between adjacent points. Use
:CALCulate:LIMit:TYPe to set the currently active limit line.
Related Command: :CALCulate:LIMit:TYPe
Front Panel Access: Shift-6 (Limit), Limit Edit, Add Vertical
:CALCulate:LIMit:X <x-parameter>
Title: Limit X Value
Description: Sets the location of the active limit line on the x-axis at the specified
location. This moves the entire limit and moves the active limit point to
the given value. <x-parameter> is defined in the current x-axis. Note
that this will change the Move Limit on the front panel to Limit if it is
currently set to Point. Use :CALCulate:LIMit:TYPe to set the currently
active limit line.
Parameter: <x-parameter>
Default Unit: Hz or for zero span in seconds
Example: To move the limit and set active point to 5 Hz:
:CALCulate:LIMit:X 5
:CALCulate:LIMit:X 5Hz
To move the limit and set active point to 500 MHz:
:CALCulate:LIMit:X 500MHz
Front Panel Access: Shift-6 (Limit), Limit Edit, Frequency