Spectrum Analyzer Commands 4-14 [:SENSe] Subsystem
S412E PM PN: 10580-00319 Rev. N 4-57
[:SENSe]:BANDwidth|BWIDth[:RESolution]:RATio <number>
[:SENSe]:BANDwidth|BWIDth[:RESolution]:RATio?
Title: Resolution Bandwidth to Span Ratio
Description: Sets the ratio of the resolution bandwidth to the span for use when the
resolution bandwidth to span coupling is enabled. Note that the front
panel interface sets the inverse ratio: the span to the resolution
bandwidth.
Parameter: <number>
Default Value: 0.01
Range: 0.00001 to 1
Related Command: :BANDwidth|BWIDth[:RESolution]:AUTO
Front Panel Access: BW, Span/RBW (note that this is the inverse ratio)
[:SENSe]:CHPower:BANDwidth|BWIDth:INTegration <freq>
[:SENSe]:CHPower:BANDwidth|BWIDth:INTegration?
Title: Channel Power Integration Bandwidth
Description: Sets the integration bandwidth for the channel power measurement.
Integration bandwidth must be less than or equal to the frequency span.
Parameter: <freq>
Default Value: 10.35 MHz
Default Unit: Hz
Related Command: :FREQuency:SPAN
Front Panel Access: Shift-4 (Measure), Channel Power, Ch Pwr Width
[:SENSe]:CHPower:STATe OFF|ON|0|1
[:SENSe]:CHPower:STATe?
Title: Channel Power Measurement State
Description: Sets the state of the channel power measurement, ON or OFF. When
using :CONFigure:CHPower, the state is automatically set to ON.
Parameter: OFF|ON|0|1
Parameter Type: <boolean>
Default Value: OFF
Related Command: :CONFigure:ACPower
Front Panel Access: Shift-4 (Measure), Channel Power, On/Off