7.2 Storage
(1) Notes on storage
When storing the MP8931A for a long period, note the followings:
•
Wipe the dust or dirt off the device before storing.
•
Avoid storing in an environment with a temperature of +70
°
C or
more, –40
°
C or less, or humidity of 80% or more.
•
Avoid places subject to direct sunlight or that are dusty.
•
Avoid places where water adhesion or corrosion by activated gas
is possible.
•
Avoid places where the equipment can be oxidized or there is ex-
treme vibration.
(2) Recommended storage conditions
In addition to the above items, the MP8931A should be stored under
the following conditions:
•
Temperature: 5 to 30
°
C
•
Humidity: 40 to 75 %
•
A place with infrequent variations in temperature and humidity
(3) Transportation
When transporting the MP8931A, use the original packaging mate-
rial. If the packaging material was not kept, pack it according to
Steps (3) and (4) below.
Re-packing procedures are as follows:
(1) Wipe the MP8931A exterior with a dry cloth.
(2) Check that no screws are loose or falling out.
(3) Provide adequate protection for projections and parts prone to
damage, and then wrap the MP8931A with a polyester sheet.
(4) Place the wrapped instrument in a cardboard box and seal it
with adhesive tape. Then, place it into a wooden box, etc. ac-
cording to the transportation distance or method.