4.2 Setting Parameters
4-7
4.2 Setting Parameters
Set measurement parameters.
Select the parameter item to be set on the main menu screen. The set-
ting procedures are shown below.
<Procedure>
1. Press [Menu] and select the parameter item to be set (ex. “Pattern”)
by using the
and
keys.
2. Press [Enter] or
>
to display the setting screen for the selected
item (ex. “Pattern”).
4.2.1 Pattern setting (Pattern)
Set pseudo-random noise pattern or fixed pattern to be measured.
Details on pseudo-random noise are shown in the block diagram below.
Select Patternon the main menu screen to perform setting.
Pattern
PN23
PN9
PN9 Reverse
PN15
PN15 Reverse
PN23
PN23 Reverse
ALL0
ALL1
1010
<Procedure>
1. Select measurement pattern by using the
and
keys on
the Pattern setting items. Press [Enter] to finalize the setting.
When this setting is performed during measurement, measurement
is restarted and the menu closes. When this setting is performed
while measurement is stopped, the measurement menu is displayed
again.
[1] PN 9: 2
9
1 (511 patterns), 1+X
5
+X
9
1
2
3
4
5
6
7
8
9
+
+
Output
Note:
Broken line indicates PN 9 Reverse (reverse direction.)