5.4 Device Message
5-23
5.4.4 MP8931A device message list
Measurement-related messages are listed in Table 5.4.4-1.
Table 5.4.4-1 Measurement-related message list
Function Message Remarks
1 Start measurement
Command START
--- Query ---
Response ---
2 Stop measurement
Command STOP
--- Query ---
Response ---
3 Measurement status
Command ---
n: Measurement status
0: Measurement
stopped
1: Measurement in
progress
Query STAT?
Response STAT n
4 Error addition
Command ERRADD
--- Query ---
Response ---
5
Error rate read-out
(Current)
Command ---
n: Error rate Query ERRRATE?
Response ERRRATE n
6
Number of er-
rors/total count
read-out (Current)
Command ---
n1: Number of errors
n2: Total count
Query ERRCNT?
Response ERRCNT n1, n2
7
Measured results
display method
Command MEASDISP n
n: Display method
0: Error Rate
1: Error/Count
Query MEASDISP?
Response MEASDISP n
8
Error rate read-out
(Last)
Command ---
n: Error rate Query ERRRATEL?
Response ERRRATEL n
9
Number of er-
rors/total count
read-out (Last)
Command ---
n1: Number of errors
n2: Total count
Query ERRCNTL?
Response ERRCNTL n1, n2
10
Elapsed measure-
ment time read-out
Command ---
n1: Hour
0 to 9999
n2: Minute
0 to 59
n3: Second
0 to 59
Query PASSTIME?
Response
PASSTIME
n1, n2, n3
11
Remaining meas-
urement time
read-out
Command ---
n1: Hour
-, 0 to 999
n2: Minute
-, 0 to 59
n3: Second
-, 0 to 59
Query LEFTTIME?
Response
LEFTTIME
n1, n2, n3

Table of Contents