3-9 WCDMA/HSDPA Signal Analyzer (Options 44, 45, and 65) Verification Performance Verification 2
3-56 PN: 10580-00177 Rev. D MS2721B MM
30. Record in the calculated results to the corresponding cells in the "Calculated ACLR Error" column of
Table A-107, and verify that they are within specification.
31. Calculate the ACLR Error at 5 MHz Offset by using the following formula:
ACLR Error at –5 MHz or +5 MHz Offset =
10 LOG
10
(10
(–45/10)
+ 10
(Measured_ACLR/10)
) - (–45) dB
32. Record in the calculated results to the corresponding cells in the "Calculated ACLR Accuracy" column of
Table A-107, and verify that they are within specification.