3-13 DVB-T/H Signal Analyzer (Options 64 and 57) Verification Performance Verification 2
3-76 PN: 10580-00177 Rev. D MS2721B MM
37. Set the MG3700A Level to 50 dBm.
38. Set the MS2721B Reference Level to 50 dBm.
39. Wait until Average (10/10) appears at the top of the display.
40. Record the Frequency Offset reading on the MS2721B into the “Frequency Error” column in Table A-125,
“DVB-T/H Signal Analyzer Frequency Accuracy for –20 dBm Reference Level, Option 64” on page A-40.
41. Record the MER Total reading on the MS2721B in the “Total MER” column in Table A-128, “DVB-T/H
Signal Analyzer Residual MER Pre Amp ON, Option 64” on page A-40.
42. Set MS2721B Pre Amp to OFF.
43. Set the MG3700A Frequency to 662 MHz and Level to 20 dBm.
44. Change the MS2721B to Channel 45, and set Reference Level to 20 dBm.
45. Repeat Step 34 through Step 41.
46. Set MS2721B Pre Amp to OFF.
47. Set the MG3700A Frequency to 854 MHz and Level to 20 dBm.
48. Change the MS2721B to Channel 69 and set Reference Level to 20 dBm.
49. Repeat Step 34 through Step 41.

Table of Contents

Other manuals for Anritsu Spectrum Master MS2721B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Anritsu Spectrum Master MS2721B and is the answer not in the manual?

Anritsu Spectrum Master MS2721B Specifications

General IconGeneral
BrandAnritsu
ModelSpectrum Master MS2721B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals