Appendix A
Decontaminating the Sample Block
Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Installation and Maintenance Guide 113
Notes
A
17. Repeat steps 8 through 11, to rinse the wells of
the sample block, and to verify that you have
eliminated the contamination.
IMPORTANT! Always use deionized water to
rinse wells after cleaning with bleach or EtOH
solution.
If contamination is present, contact Applied
Biosystems technical support (see page xiii).
18. Ensure that the heated cover assembly door is
completely closed. If it is not, an error message is
displayed.
10% bleach solution