Chapter 5 Performing the Background Calibration and Optical Calibration
Performing the Optical Calibration for the 7500/7500 Fast System
62 Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Installation and Maintenance Guide
Notes
Performing the Optical Calibration
1. In the SDS software:
a. Select the Instrument tab.
b. Click .
The instrument begins the calibration.
Note: Before starting the calibration, the
instrument may pause (up to 10 minutes) to
allow the heated cover to reach the correct
temperature.
2. When the run is complete, click .
a
b
Click

Table of Contents