HA-8180V 3-39
b On the standby screen, select , <2 Reagent replacement
menu> and <6 Edit reagent information>.
The current information will appear.
Hyphens (---) appear where no information has been set.
c Select a reagent.
button: Moves the cursor to the next setup field.
d Below <Code>, enter the 10-digit reagent code.
For eluents and hemolysis washing solution, the code is written
on the pack or bottle labels.
For calibrator, the code is written on the Standard value list.
If your entry is accepted, the lot number and expiration date will
be updated (or appear).
e To set information on another reagent, repeat from step b.
f Press .
This saves your entries and will return you to the [Reagent replacement menu] screen.
g To return to the standby screen, press twice.
Edit reagent information
(2600) 10-06-2010 10:00
Code Lot No. Expiry
Elu. 80A <----------> ------- -------
Elu. 80B <----------> ------- -------
Elu. 80CV<----------> ------- -------
Wash 80H <----------> ------- -------
Calib. <----------> ------- -------
PrevItem BS OK Cancel
Menu
REFERENCE:
For eluents and hemolysis washing solution, the optional hand-held barcode reader can be used to read the
reagent codes.
To delete the current entry, press the hyphen button. Then hyphens (---) appear.
NOTE:
If an error message appears:
The reagent information is invalid. See the table below to determine the cause. Press to return to
step b and enter the correct code. Invalid entries will then appear as hyphens (---).
When setting information for two or more reagents at a time:
You can go to step g once all of your entries are valid. If invalid information was entered for two or more
reagents, repeat steps b to f until no more error messages appear.
OK
OK
Go back
Error message Description
Reagent type in reagent information does not
match the selected reagent type.
You entered the reagent code of another reagent. For example, you
entered the reagent code of Eluent 80B in <Elu. 80A>.
Manufacturing date in reagent information is
not consistent with the current date.
The manufacturing date in the reagent information is after the
current date.
Expiration date in reagent information has
passed.
The expiration date in the reagent information is before the current
date.
3.8 Reagent Information Settings

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Arkray ADAMS A1C HA-8180V and is the answer not in the manual?

Arkray ADAMS A1C HA-8180V Specifications

General IconGeneral
BrandArkray
ModelADAMS A1C HA-8180V
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals