HA-8180V 4-5
b Wipe any liquid from the nozzle with a new piece of gauze.
c Insert the nozzle of the bottle cap into the new pack and tighten
the cap securely.
d Check that the cap of the eluent pack is tightened securely.
e Check the type of the new eluent pack and place the pack in the specified position on the bottle tray.
f Shape the eluent pack into a smooth standing position.
5 Reset the remaining reagent graphic for the eluent.
a Press .
• This resets the remaining reagent graphic for the eluent.
• The [Reagent replacement menu] screen will appear again.
Remove any lint if it is stuck to the nozzle. Lint may clog the tube.
Attach the bottle cap with nozzle to the pack somewhere
other than above the instrument. Liquid may spill and
damage the instrument.
If the cap is loose, liquid condensation may rise due to
evaporation, resulting in inaccurate measurement results.
Set the pack between the applicable pack supporters.
Eluent may not be fully aspirated if the pack is bent over or collapses during measurement.
If an error message appears:
The reagent information is invalid. Press to return to step 2. Check the type, expiration date and
manufacturing date of the eluent, and enter the correct reagent code. To replace with another new eluent pack,
repeat steps 2 to 4.
OK
4.2 Replacement of Consumables