HA-8180V 4-4
Chapter 4 Maintenance
a Below <Code>, enter the 10-digit reagent code that is written on
the label of a new eluent pack.
If your entry is accepted, the lot number and expiration date will
be updated.
3 Remove the used eluent pack.
a Lay out some gauze near the instrument.
b Remove the used eluent pack from the bottle tray.
c Remove the bottle cap with nozzle from the pack.
Place the nozzle on the gauze.
4 Place the new eluent pack.
a Hold the new eluent pack by the hard plastic neck and remove the
cap from the pack.
Eluent A
(2100) 10-06-2010 10:00
Code Lot No. Expiry
<1234567890> 0A1101 2011-01
Replace with new pack.
Hold pack by hard plastic neck.
Holding by the bag
may spill contents.
BS Finish Go back
Reagent code on
the pack label
NOTE:
If <Lot No.> and <Expiry> are not updated and still contain
hyphens (---):
Your entry may be incorrect. Carefully check the reagent
code and enter it again.
REFERENCE:
You can use the optional hand-held barcode reader to read the
reagent code.
Hold here.
NOTE:
Do not hold the eluent pack by the soft aluminium bag.
Eluent may spill and damage the instrument.
REFERENCE:
Keep the cap in the accessory case. This cap should be reused when transporting the instrument, or if the
instrument is not to be used for extended periods of time.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Arkray ADAMS A1C HA-8180V and is the answer not in the manual?

Arkray ADAMS A1C HA-8180V Specifications

General IconGeneral
BrandArkray
ModelADAMS A1C HA-8180V
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals