EasyManuals Logo

Arkray ADAMS A1C HA-8180V User Manual

Arkray ADAMS A1C HA-8180V
246 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #174 background imageLoading...
Page #174 background image
HA-8180V 4-27
2 Prepare the washing solution for tubes.
a Add 7 mL of washing solution for tubes to each of two sample
tubes.
b Set the sample tubes in ports 1 and 2 of the washing rack.
c Load the washing rack onto the sampler.
See step 4 in “2.5.1. Preparing Samples” on page 2-21.
3 Wash the tubes.
a Press or .
Tube washing will start.
4 Turn off the power.
The message “Tube washing ended. Turn off power.” will appear when tube washing is complete.
a Check that the washing rack is not moving and remove the rack from the rack unloading side of the sampler.
b Press the standby switch to turn off the power.
Washing solution
for tubes (7 mL)
12345678910
NOTE:
Be sure to use the washing rack. Use of other racks may
damage the instrument or seriously degrade the column,
requiring its replacement.
REFERENCE:
To have the instrument enter sleep mode automatically after tube washing:
Press . If the startup timer is set to [Use], select the day the next time the instrument starts up (see
“3.3.3. Setting the Timer” on page 3-9).
NOTE:
If the timer was set by pressing , do not turn off the power by pressing the standby switch in step b
below. Turning off the power with the standby switch will cancel the timer.
REFERENCE:
The washing solution for tubes will remain in the instrument and will be drained the next time the instrument starts
up.
Start
Timer
Timer
4.3 Washing and Cleaning

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Arkray ADAMS A1C HA-8180V and is the answer not in the manual?

Arkray ADAMS A1C HA-8180V Specifications

General IconGeneral
BrandArkray
ModelADAMS A1C HA-8180V
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals