11. Specification and accessories
page 87 / 96 Issue date 18.05.2021, rev. 1.0
11. Specification and accessories
Technical data
Classifications:
Medical device class: IIa
Classification rule: 9
(according to MDD 93/42 / EEC and REGULATION (EU) 2017/745 OF THE EUROPEAN PARLIAMENT AND
OF THE COUNCIL of 5 April 2017)
Electrical safety class: II
Applied part type: BF
Degree of protection provided by unit enclosure: IP20
Mode of operation:
The unit is intended for continuous operation.
Treatment parameters:
Described in chapter 8 and
Nominal load resistance: 500 - 750 Ω
Accuracy of operation parameters:
Output current and voltage amplitude: ±20% for the load resistance: 500 Ω
Microcurrents calibration: for resistance 22 kΩ
Pulse repetition frequency: ±20%
Pulse duration: ±20%
Accuracy of times of individual phases for AM and FM: ±20%
Programs and sequences:
Preset programs: 71
Preset sequences: 44
Total 115
User defined:
Programs: 50
Sequences: 10
Treatment timer:
Ranges and resolutions: defined in chapter 8
Time accuracy: 10%
General:
Mains supply: 100-240 V; 50/60 Hz
PhysioGo.Lite Electro controller supply: 24VDC; 2,5A
Mains fuses: size 5x20mm, T3,15L250V; 3,15 A, 250 V
Type of memory backup battery: CR2032
Unit weight: max. 3 kg
Unit dimensions (WxDxH): 25x27x16,5cm
Switched-mode power supply, type HPU63B-108 by Sinpro:
Mains supply – input: 100-240 VAC; 1.62-0.72A; 47-63 Hz
Output: max. 24VDC; 2.62A
Weight: max. 0.38 kg
Dimensions (WxDxH): 13.2x5.6x3.7 cm