8. Definitions and parameters
User ManualPhysioGo.Lite Electro page 58 / 96
Symbol Description
Symbol on the
treatment
screen
Available parameters
ALT
Automatic polarization switch in the half
performed treatment.
Amplifier
operation mode
-
CC stabilized output current
CV stabilized output voltage
Maximum
amplitude
-
Unipolar
current
0,330 mA in CC mode
Regulation step:
0,1 mA in the range of 0,3-10 mA
0,5 mA in the range of 10-30 mA
1,5100 V in CV mode, max. 30 mA
Regulation step:
0,5 V in the range of 1,5-100 V
Bipolar
current
0,5 – 100 mA in CC mode
Regulation step:
0,1 mA in the range of 0,5-10 mA
0,5 mA in the range of 10-100 mA
1,5100 V in CV mode, max. 100 mA
Regulation step:
0,5 V in the range of 1,5-100 V
Leduc’s current
Another are Leduc’s current is a special case of rectangular pulse currents with pulse duration 1 ms and pause
duration 9 ms.
Parameters description:
Symbol Description
Symbol on
the treatment
screen
Available parameters
Treatment time - 1 60 minutes, 1 minute step
Shape of the
current
-
Rectangular pulse current according to
Leduc
Continuous or
interrupted
shape of the
current
c_shape
Continuous
Pulse frequency is 4 kHz
Duty factor 80%
Pulse frequency is 8 kHz
Duty factor 90%
Pulse duration - Constant, 1 ms
Pause duration - Constant, 9 ms
Basic pulse
frequency
- Constant, 100 Hz
Training
program
PT
No training program, non-editable
parameter
Polarization NOR
For such polarization setting red plug is a
positive electrode, and black plug is a
negative electrode.