8. Definitions and parameters
User Manual – PhysioGo.Lite Electro page 58 / 96
Symbol Description
treatment
Available parameters
ALT
Automatic polarization switch in the half
performed treatment.
Amplifier
operation mode
-
CC – stabilized output current
CV – stabilized output voltage
Maximum
amplitude
-
Unipolar
current
Regulation step:
• 0,1 mA in the range of 0,3-10 mA
• 0,5 mA in the range of 10-30 mA
1,5 – 100 V in CV mode, max. 30 mA
Regulation step:
•
0,5 V in the range of 1,5-100 V
Bipolar
current
Regulation step:
• 0,1 mA in the range of 0,5-10 mA
• 0,5 mA in the range of 10-100 mA
1,5 – 100 V in CV mode, max. 100 mA
Regulation step:
• 0,5 V in the range of 1,5-100 V
Leduc’s current
Another are Leduc’s current is a special case of rectangular pulse currents with pulse duration 1 ms and pause
duration 9 ms.
Parameters description:
Symbol Description
the treatment
Available parameters
Treatment time - 1 – 60 minutes, 1 minute step
-
Rectangular pulse current according to
Leduc
Continuous or
interrupted
shape of the
current
c_shape
Continuous
• Pulse frequency is 4 kHz
•
• Pulse frequency is 8 kHz
• Duty factor 90%
Pulse duration - Constant, 1 ms
Pause duration - Constant, 9 ms
- Constant, 100 Hz
PT
No training program, non-editable
parameter
Polarization NOR
For such polarization setting red plug is a
positive electrode, and black plug is a