10. Maintenance, cleaning, disinfection
User Manual – PhysioGo.Lite Electro page 86 / 96
Symptoms Undertaking action
Unit equipped with battery
module – the device does not
Connect the mains supply. The battery may be discharged.
To start the operation, please hold on for at least 5 seconds the STANDBY
key.
The battery discharges quickly
Contact your authorized service for battery replacement. If the battery
module has to be dismantled, a stabilizing cartridge should be installed.
If you change the battery yourself, follow the information included in 5.3
Date and time settings cancel
If I16 error is shown on the display, it means that the backup battery is
discharged. Its exchange should be directed to an authorized service.
Type of memory backup battery is a CR2032.
Fuse replacement
NOTE:
Before proceeding to the further described operations isolate the unit and switch mode power supply from
the mains supply!
In case of burnt the fuse, it must be replaced. Fuse parameters are given in chapter ”Specification
and accessories” and on the name plate.
To replace the fuse:
Step Description
Switch the power switch to the "0" position.
2.
Disconnect the mains cable from the mains socket. Disconnect the power supply cable from the
device socket marked with symbol .
With flat screwdriver unscrew the fuse socket until the moment of its slipping from the socket.
4.
Remove the socket with your fingers, replace the fuse, install them in the socket again and screw
firmly.
5.
Connect the power supply cord to the device socket. Then connect the mains cable to the mains
socket.
6. Switch the power switch on and start the device. Check the device operation.