8. Definitions and parameters
User Manual – PhysioGo.Lite Electro page 50 / 96
Symbol Description
treatment
Available parameters
Contraction phase 10 s
Fall time 2 s
Contraction phase 15 s
Fall time 3 s
Contraction phase 15 s
Fall time 3 s
operation
-
CC – stabilized output current
CV – stabilized output voltage
Maximum
amplitude
-
Regulation step:
• 0,5 mA in the range of 0,5-140 mA
1,5 – 100 V in CV mode, max. 140 mA
Regulation step:
•
0,5 V in the range of 1,5-100 V
Medium frequency currents
The idea of using medium frequency currents is similar to the use of Russian stimulation, but with other carrier
frequencies available. This enables the choice of stimulation depth.
Parameters description:
Symbol Description
treatment
Available parameters
Treatment
time
- 1 – 60 minutes, 1 minute step
Carrier
frequency
c_freq Possible settings: 2 kHz, 4 kHz, 8 kHz
Basic
frequency
f_bas Regulation in the range of 5 Hz – 100 Hz (variable step)
Training
program
PT
No training program
Contraction phase 2 s
Fall time 1 s
Contraction phase 3 s
Fall time 1 s