8. Definitions and parameters
User Manual – PhysioGo.Lite Electro page 66 / 96
Symbol Description
the treatment
Available parameters
Pause duration t_idle
• 80 ms – IG30-UNI, IG30-BI
• 5 ms – IG50-UNI, IG50-BI, IG100-UNI, IG100-BI,
Pulse edge rise
time
t_r
• 30 ms – IG30-UNI, IG30-BI
• 300 µs – IG50-UNI, IG50-BI, IG100-UNI, IG100-BI,
Pulse edge fall
time
t_f
• 10 ms – IG30-UNI, IG30-BI
• 100 µs – IG50-UNI, IG50-BI, IG100-UNI, IG100-BI,
Training program PT
No training program, non-editable
parameter
Contraction phase 15 s
Fall time 10 s
Rest phase 100 s
Contraction phase 65 s
Fall time 10 s
Rest phase 150 s
Contraction phase 115 s
Fall time 10 s
Rest phase 200 s
Maximum
amplitude
-
Unipolar
current
Regulation step:
• 0,1 mA in the range of 0,3-10 mA
• 0,5 mA in the range of 10-80 mA
1,5 – 100 V in CV mode, max. 60 mA
Regulation step:
•
0,5 V in the range of 1,5 -100 V
Bipolar
current
Regulation step:
• 0,1 mA in the range of 0,5-10 mA
• 0,5 mA in the range of 10-80 mA
1,5 – 100 V in CV mode, max. 60 mA
Regulation step:
•
0,5 V in the range of 1,5-100 V
EMS
EMS i.e. electrostimulation of properly innervated muscles is based on generating a contraction of selected
muscles with the appropriately selected rectangular symmetric impulses. Its aim is to maintain the contraction
ability of the properly innervated muscle, to prevent muscle atrophy and to restore muscle strength and mass.