8. Definitions and parameters
page 67 / 96 Issue date 18.05.2021, rev. 1.0
Parameters description:
Symbol Description
treatment
Available parameters
Treatment time - 1 – 60 minutes, 1 minute step
-
Symmetric
Pulse duration t_imp
Possible settings: 25 µs, 50 µs, 75 µs, 100 µs, 150 µs, 200
µs, 250 µs, 300 µs, 400 µs, 500 µs
Basic frequency f_bas Regulation in the range of 1 Hz – 250 Hz (variable step)
Frequency
spectrum
f_spec Regulation in the range of 0 Hz – 250 Hz (variable step)
Frequency
modulation
program
FM
Frequency modulation program is switched
off
Hold time of maximum frequency 2 s
Frequency fall time 1 s
Hold time of basic frequency 3 s
Frequency rise time 3 s
Frequency fall time 3 s
Hold time of maximum frequency 3 s
Frequency fall time 3 s
Hold time of basic frequency 3 s
Hold time of maximum frequency 6 s
Frequency fall time 3 s
Hold time of basic frequency 6 s
Hold time of maximum frequency 6 s
Frequency fall time 3 s
Hold time of basic frequency 12 s
Hold time of maximum frequency 12 s
Frequency fall time 3 s
Hold time of basic frequency 6 s
Frequency rise time 6 s
Frequency fall time 6 s
Hold time of maximum frequency 6 s
Frequency fall time 6 s
Hold time of basic frequency 6 s
Frequency rise time 12 s
Frequency fall time 12 s
Amplitude
modulation
program
AM
Amplitude modulation program is switched
off
Contraction phase 3 s
Fall time 1 s
Contraction phase 5 s
Fall time 1 s