8. Definitions and parameters
User ManualPhysioGo.Lite Electro page 52 / 96
SP-TENS pulse current
On the basis of classical TENS pulse currents, SP-TENS currents were created, intended for spastic paralysis
treatments. Stimulation may be performed by using one- or two-channels simultaneously. While two-channel
operation, during the stimulation phase in one channel, in the second channel the rest phase occurs, then there
is a change.
Parameters description:
Symbol Description
Symbol on
the treatment
screen
Available parameters
Treatment time - 1 60 minutes, 1 minute step
SP-TENS current -
One-channel
Sequence, available only in
TENS pulses type in
A and B circuits
SYM
Symmetric
ASYM
Asymmetric
TENS pulses
duration time in A
and B circuits
t_imp
Possible settings:: 25 µs, 50 µs, 75 µs, 100 µs, 125 µs, 150
µs, 175µs, 200 µs, 250 µs, 300 µs
Basic frequency in A
and B circuits
f_bas Regulation in the range of 30 Hz 100 Hz (step 10 Hz)
Training program PT
Stimulation phase 2 s
Rest phase 2 s
Stimulation phase 2 s
Rest phase 4 s
Stimulation phase 2 s
Rest phase 6 s
Stimulation phase 2 s
Rest phase 10 s
Rise time 0,5 s
Stimulation phase 3 s
Fall time 0,5 s
Rest phase 4 s
Rise time 0,5 s
Stimulation phase 3 s
Fall time 0,5 s
Rest phase 8 s
Rise time 1 s
Stimulation phase 4 s
Fall time 1 s
Rest phase 6 s
Rise time 1 s
Stimulation phase 4 s
Fall time 1 s
Rest phase 10 s

Table of Contents