5. Unit description
User Manual – PhysioGo.Lite Electro page 22 / 96
5.6.2 Electrodes condition control
This function is intended to inform about deteriorating condition of electrodes, so as to give the operator time
to stock up on new accessories.
The condition of the electrodes is signaled by the color of the electrotherapy indicator on the treatment screen
(Figure 5.7):
• white – electrodes in good condition,
• flashing yellow – worn electrodes.
If wear of the electrodes is detected, the device will display an appropriate message at the end of the treatment
– see Figure 5.6 ("Poor contact of electrodes").
Figure 5.7. Signalization of worn electrodes
5.6.3 Current accuracy control in CC mode
While performing the electrotherapy treatment in stabilized output current mode (CC), the unit controls
the accuracy of current intensity. In the case when the difference between the setting and output value is higher
than 20%, the treatment will be interrupted and the message will appear on the display.
5.6.4 Overcurrent in CV mode
While performing the electrotherapy treatment in stabilized output voltage mode (CV), the unit controls whether
the maximum current value is not beyond the limits. If the voltage setting for current value is higher than the
allowed limit (specified for the stabilized output current in CC mode), the treatment will be interrupted and the
message will appear on the display.
5.6.5 Information on the features of generated signal
This function is intended to inform about:
• classification of currents due to their polarity,
• warning threshold of the current value that requires an operator's special attention in the generation phase
(e.g. training programs, modulations, pulse current).
Detailed information and symbols are provided – see Table 5.6.