BD BACTEC™ FX Instrument User’s Manual
162
25
Drawer X:
Measurement
System offline.
Agitation failure has set the
measurement system to
offline after three consecutive
failures to stop at the read
position.
Reboot instrument.
Stations are marked as Unusable.
Refer to Section
7.3.2 for additional
information on dealing with Unusable
stations.
Move vials in the indicated stations
within 10 minutes of the set time of this
alert to prevent them from becoming
Affected Vials.
26
Drawer open too
long.
The drawer has been open
for longer than 10 minutes.
Close the drawer. Allow it to remain
closed for at least 30 minutes.
NOTE: If drawer is not closed within 40
minutes of the time that it is opened, all
vials in drawer are marked as Affected
Vials. Refer to Section
7.3.3 for
instructions on Affected Vials.
30
The instrument has
lost connectivity to
the server
database.
The instrument has lost
communications with the
BD
EpiCenter master
database.
Instrument enters a degraded mode of
operation. See Section
4.12.3 for
information on degraded operations.
31
Database Object
Reinitialized.
An individual reading or a
reading collection is
corrupted.
A sector on the flash drive is corrupted
or a bad checksum is encountered on
a reading object or collection. One or
more readings have been lost.
Message is informational. If four
consecutive readings become
corrupted, then a reading gap will
occur and the vial will automatically
become affected.
32
Barcode Reader n:
Cannot determine
type.
Instrument cannot
communicate with barcode
reader to determine the
barcode reader’s type.
Communication attempt continues
every two minutes until successful
communication with the barcode
reader is established.
33
Program download
failure.
Set when a microprocessor
download fails to complete
successfully.
Reboot instrument.
34
Watchdog Timeout
Failure, Notify
Becton Dickinson.
An error occurred in the main
computer software.
Reboot the main computer using the
Maintenance > Utilities > Reboot
function.
Error
No.
Message Possible Cause(s) Corrective Action(s)