7 – Troubleshooting
165
49
Drawer X Rows A &
B are Offline.
Remove any vials
in unusable
stations. Consult
Manual.
When the Drawer Control
Board fails to communicate
with the Row Board that
controls rows A and B, the
Row Board is marked offline.
Reboot instrument.
Stations are marked as Unusable.
Refer to Section
7.3.2 for additional
information on dealing with Unusable
stations.
Move vials in the indicated stations
within 40 minutes of the set time of this
alert to prevent them from becoming
Affected Vials.
50
Drawer X Rows C &
D are Offline.
Remove any vials
in unusable
stations. Consult
Manual.
When the Drawer Control
Board fails to communicate
with the Row Board that
controls rows C and D, the
Row Board is marked offline.
Reboot instrument.
Stations are marked as Unusable.
Refer to Section
7.3.2 for additional
information on dealing with Unusable
stations.
Move vials in the indicated stations
within 40 minutes of the set time of this
alert to prevent them from becoming
Affected Vials.
51
Drawer X Rows E &
F are Offline.
Remove any vials
in unusable
stations. Consult
Manual.
When the Drawer Control
Board fails to communicate
with the Row Board that
controls rows E and F, the
Row Board is marked offline.
Reboot instrument.
Stations are marked as Unusable.
Refer to Section
7.3.2 for additional
information on dealing with Unusable
stations.
Move vials in the indicated stations
within 40 minutes of the set time of this
alert to prevent them from becoming
Affected Vials.
52
Drawer X Rows G
& H are Offline.
Remove any vials
in unusable
stations. Consult
Manual.
When the Drawer Control
Board fails to communicate
with the Row Board that
controls rows G and H, the
Row Board is marked offline.
Reboot instrument.
Stations are marked as Unusable.
Refer to Section
7.3.2 for additional
information on dealing with Unusable
stations.
Move vials in the indicated stations
within 40 minutes of the set time of this
alert to prevent them from becoming
Affected Vials.
53
Drawer X Rows J &
K are Offline.
Remove any vials
in unusable
stations. Consult
Manual.
When the Drawer Control
Board fails to communicate
with the Row Board that
controls rows J and K, the
Row Board is marked offline.
Reboot instrument.
Stations are marked as Unusable.
Refer to Section
7.3.2 for additional
information on dealing with Unusable
stations.
Move vials in the indicated stations
within 40 minutes of the set time of this
alert to prevent them from becoming
Affected Vials.
Error
No.
Message Possible Cause(s) Corrective Action(s)