BD BACTEC™ FX Instrument User’s Manual
166
Barcode Messages
BC01
Invalid medium
type. Reenter
barcodes.
A vial sequence number was
scanned or entered and the
media type is not defined in
the instrument.
Make sure the correct vial barcode or
replacement vial barcode is scanned;
only original BD vial sequence or
BD-supplied replacement barcodes
can be used for sequence numbers.
If vial sequence number is entered
manually, be careful to enter it
correctly.
Select OK to remove the message
box.
BC03
Invalid sequence.
Reenter barcodes.
The vial sequence number
was entered or scanned that
does not meet the defined
parameters (e.g., it is too
long, too short, has incorrect
digits).
Make sure the correct vial barcode or
replacement vial barcode is scanned;
only original BD vial sequence or
BD-supplied replacement barcodes
can be used for sequence numbers.
If vial sequence number is entered
manually, be careful to enter it
correctly.
Select OK to remove the message
box.
BC05
Invalid accession.
Reenter barcodes.
An accession number was
entered that does not meet
the defined parameters. It
could contain illegal
characters such as:
* ? [ ] ! # |
or it could have too many
digits. You could also have
scanned a sequence already,
and then scan another
sequence when the
instrument is expecting an
accession barcode scan.
Enter a valid accession number, up to
20 characters that does not contain the
following characters:
* ? [ ] ! # |
Select OK to remove the message
box.
Error
No.
Message Possible Cause(s) Corrective Action(s)