7 – Troubleshooting
167
Culture Screen Messages
CS01
Patient ID not
found.
You entered a patient ID that
is not in the database.
The value you entered is
shown at the top of the
message box.
Make sure Patient ID is entered
correctly and completely. You cannot
enter a partial ID to recall patient
information.
CS02
Patient name not
found.
You entered a patient name
or portion of a patient name
that is not in the database.
The value you entered is
shown at the top of the
message box.
Try entering only the first portion of the
name if you tried entering the whole
name.
CS03
Too many patients
found. Please
refine search.
A Patient Name search
matched more than 50 entries
in the database.
Enter more characters to narrow the
search results.
CS15
Disassociate
sequence from
accession?
You have pressed the
Disassoc(iate) button on the
Culture – Vial display.
The sequence and accession
are shown at the top of the
message box.
Select Yes to confirm the
disassociation.
Select No to cancel the disassociation.
CS21
One or more
reports have been
canceled.
Appears if you select Cancel
to a print request.
Message is informational.
CS22
Report has been
sent to the printer.
On the Plot display, the Print
button was pressed. Also
generated when you respond
Yes to message
WE35, or
when printing is requested for
non-oversize reports.
Message is informational.
Select OK to remove the message box
and print the report.
CS23
Disassociate
specimen from
patient?
You have pressed the
Disassoc(iate) button on the
Culture – Patient display.
The sequence and accession
is shown at the top of the
message box.
Select Yes to confirm the
disassociation.
Select No to cancel the disassociation.
CS24
Sequence scanned
already associated
with accession.
Must disassociate
on Vial tab before
reassociating.
Message occurs if you scan
any non-sequence barcode
(other than one matching the
displayed accession) or a
sequence that is already
associated to a different
accession number.
A vial can only be attached to one
accession number.
Error
No.
Message Possible Cause(s) Corrective Action(s)