BD BACTEC™ FX Instrument User’s Manual
168
ID Anonymous Messages
ID01
Positive-
Anonymous pulled.
Scan sequence and
touch <Save> to ID
and remove.
A positive anonymous vial
was pulled from a station.
The station is shown at the
top of the message box.
Scan the vial sequence barcode.
Select Save to save the identification if
you are keeping the vial out of the
instrument.
If you are returning the vial, place it in
the FLASHING GREEN station and do
not select Save.
ID02
The instrument has
lost connectivity to
the server
database. Vial
cannot be identified
now and must be
returned
anonymously to
maintain readings.
Touch Cancel to
discard all
readings.
An anonymous vial was
pulled from a station in an
instrument that is in degraded
mode.
Place the vial back into the same
station to continue testing the vial
anonymously.
Or select Cancel or close the drawer
to discard all readings.
ID05
Vial pending
identification. All
readings will be
lost.
This is displayed when
<Discard> is selected on the
Identify Anonymous Screen.
Select OK to discard the vial’s
readings.
Select Cancel to cancel the Discard
operation.
ID09
Removed vial is
anonymous.
Identify?
An anonymous vial is pulled
when a display other than ID
Anonymous is displayed.
Select Yes to identify the anonymous
vial. The ID Anonymous display
appears.
Select No if you do not want to identify
the anonymous vial. Additional
message(s) provide further
instructions.
ID10
Vial has been out of
the instrument too
long. The vial's
protocol will be
restarted and it
should be
subcultured.
Consult Manual.
<OK>.
The vial barcode sequence
scanned belonged to a known
vial which was removed from
the instrument more than 5
hours (reentry window) ago.
If the vial is placed back in the
instrument, it is treated as a new vial.
If this occurs during ID Anonymous
activity, the vial maintains all the test
readings and information associated to
the anonymous vial, but the previous
sequence information is discarded.
ID12
Re-insert vial to
continue measuring
anonymously or
touch Cancel to
discard all
readings.
Occurs if you respond No to
message
ID09, or if you
select Return on the ID
Anonymous display.
Vial continues as anonymous if you
place it back in the station. Previous
test readings are retained and testing
continues.
If you select Cancel in response to the
message, the vial becomes a newly
entered anonymous vial.
Error
No.
Message Possible Cause(s) Corrective Action(s)