7 – Troubleshooting
169
ID13
Vial pending
identification.
Discard all readings
and exit workflow?
Occurs if you select Exit on
the ID Anonymous display,
with information related to a
pulled vial on the screen.
Select Yes to exit the ID Anonymous
display.
All readings to date for the vial are
discarded.
Select No to cancel the Exit operation
and continue identifying anonymous
vials.
ID14
Vial cannot be
identified with this
sequence.
Duplicate sequence
exists. Consult
Manual.
The instrument has
determined that the sequence
number you just scanned
belongs to a different vial.
A vial swap has occurred.
For optimal recovery, both vials should
be subcultured.
To reenter vials, use the Vial Entry
activity.