BD BACTEC™ FX Instrument User’s Manual
164
38
Drawer X: Contains
a vial with a reading
gap. Consult
manual.
The reading gap evaluator
determines when a vial has a
reading gap greater than 40
minutes or the algorithms
have not processed readings
for 40 minutes. This alert is
reported each time a different
vial with a reading gap is
detected in that drawer.
Any vials in the affected row(s) are
marked as Affected Vials. Refer to
Section
7.3.3 for instructions on
Affected Vials.
39
Drawer X: A
temporary agitation
failure has
occurred.
Agitation failure caused by
drawer not being fully closed.
Open and close the drawer, and then
power the instrument OFF and back
ON. Contact BD.
40
Reminder – The
drawer has been
open too long.
Caused when the drawer is
still open every 5 minutes
after Alert
26 has been
reported and acknowledged.
Close the drawer.
NOTE: If drawer is not closed within 40
minutes of the time that it is opened, all
vials in drawer are marked as Affected
Vials. Refer to Section
7.3.3 for
instructions on Affected Vials.
41
NTP server
unavailable: clocks
may not be
synchronized.
Set when the network client
cannot get a response to a
time synchronization request.
Make sure all network cables are
plugged in.
44
BD BACTEC FX
and BD EpiCenter
times are not
synchronized.
The instrument has
determined that its time is not
synchronized with the
BD
EpiCenter Time Service.
Make sure all network cables are
plugged in.
46
Database recovery
file invalid.
This will only cause
(potentially) the last
transaction before the power
failure to be rolled back. A
new recovery file will be
created.
Message is informational only. No
action necessary.
47
Drawer X: Agitation
Re-Started
Agitation speed is outside of
normal range, or failed to stop
at a sensor or see a sensor
for 10 continuous seconds.
A Drawer Open,
Measurement scan or Power
failure resets four times
consecutively.
Alert is reported on System Alerts
display and report only.
Alert is cleared when agitation speed
returns within range.
If this message recurs frequently,
contact BD for service.
Error
No.
Message Possible Cause(s) Corrective Action(s)