Chapter 4: DNA Analysis 83
Figure 4-4 Defining singlet and doublet populations
9 In the Acquisition Dashboard, click Record Data.
10 When recording finishes, note the CV of the Singlet population.
•If the CV is ≤3%, continue to step 11.
• If the CV is >3%, restart the acquisition. Decrease the flow rate with
the SAMPLE FINE ADJ knob until the CV is ≤3%, and then re-record
the data.
If the CV does not improve, see Troubleshooting on page 113.
CV <3%