42 BD LSR II User’s Guide
Changing Optical Filters or Mirrors
1 Lift the appropriate instrument cover.
• The octagon array is located under the right instrument cover.
• The three trigon arrays are located under the left instrument cover.
; Tip To open the left instrument cover, you must first open the right cover
and the side door.
2 Remove the appropriate filter holder or mirror holder.
3 Replace it with the new filter holder or mirror holder.
; Tip The filters and mirrors fit easily into the optical holders in only one
way.
4 Close the instrument cover(s).
Custom Configurations
See Custom Configurations in Appendix E for specifications of some common
custom filter and mirror configurations. The custom filters and mirrors used in
these configurations are included with your spares kit.
Follow the precautions outlined in Laser Safety on page xvii while changing
optical filters or mirrors.
LSR2.book Page 42 Tuesday, April 25, 2006 3:34 PM