46 BD LSR II User’s Guide
Preparing the Waste Container
Figure 2-4 Waste container
1 Disconnect the orange waste tubing and the black level sensor line from the
waste container.
Keep the lid on the waste container until you are ready to empty it.
2 Empty the waste container.
All biological specimens and materials coming into contact with them are
considered biohazardous. Handle as if capable of transmitting infection.
Dispose of waste using proper precautions and in accordance with local
regulations. Never pipette by mouth. Wear suitable protective clothing,
eyewear, and gloves.
The waste container contents might be biohazardous. Treat contents with
bleach (10% of total volume). Dispose of waste using proper precautions
and in accordance with local regulations. Wear suitable protective clothing,
eyewear, and gloves.
waste tubing
(from cytometer)
level sensor line
bracket
LSR2.book Page 46 Tuesday, April 25, 2006 3:34 PM