110 BD LSR II User’s Guide
Changing the Bal Seal
The sample injection tube Bal seal is a Teflon ring that forms a seal with the
sample tube and ensures proper tube pressurization. Over time, this seal becomes
worn or cracked and requires replacement. Replacement is necessary if a proper
seal is not formed when a sample tube is installed on the SIP.
; Tip Indications that a proper seal has not formed include:
the tube will not stay on the SIP without the tube support arm
the tube is installed, RUN is pressed on the cytometer, and the RUN button
is orange (not green)
To replace the Bal seal:
1 Remove the outer droplet sleeve from the sample injection tube by turning
the retainer counterclockwise.
Figure 6-6 Removing the outer sleeve
; Tip Work carefully—the outer sleeve can fall out as you loosen the
retainer.
Bal seal
LSR2.book Page 110 Tuesday, April 25, 2006 3:34 PM