9-12
TotalAlert Innity™ HTM Medical Gas Notication System
4107 9016 60.01
Figure 280
9.4 Combination Alarm congura-
tion - Website
1. Accessing the Web Page.
Start a web browser such as
Chrome,FireFox,andSafari (Browser
needs to be able to interpret HTML5)
2. Enter the Device’s IP address in the
browser’s address bar:
Example: http://169.254.100.100
(IP address of the computer needs to be in
the range with this device)
3. After you enter the IP address, the
alarm’s home page will be
displayed. See Figure 281
NOTE:
To learn the IP address, touch the MISCEL-
LANEOUS tab on the configuration screen
then touch the WIRED ETHERNET tab.
Figure 281
Login to Setup Pages
1. Click Login and Setup.
2. The web browser will request a user-
name and password (Figure 282). The
factory defaults are:
Username: TotalAlert
Password: 121212
NOTE:
If the on-screen password has been changed
from the default password, use the newly
assigned password.
Figure 282
3. The left side bar will now contain the “La-
bel Setup” link. (See Figure 283)
4. Click on Label Setup to enter the main
label screen for non-latin language input.