EasyManua.ls Logo

BeaconMedaes TotalAlert Infinity - Setting Up Combination Alarms

BeaconMedaes TotalAlert Infinity
122 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4-20
TotalAlert Innity™ HTM Medical Gas Notication System
4107 9016 60.01
Figure 70: Combination Alarm Main Screen
NOTE:
See the end of this section for information on the
Miscellaneous Tab. See Section 4.6 for information
on the Active Alarm Opt Tab.
4.5 Setting up a Combination
Alarm
The TotalAlert Innity
combination alarm combines
the functionality of a Area alarm and a central alarm. It
monitors two things:
Gases and their pressures.
Source equipment.
4.5.1 Initial Setup of an Uncongured
Combination Alarm
To begin setup of the alarm, rst pull the battery tab
located on the back of the front panel to activate
battery backup of the alarm.
4.4.3 Maintenance Mode Setup
To ensure the audible warning does not sound while
setting up the alarm, the MAINTENANCE MUTE function
can be activated.
Touch the MISCELLANEOUS tab in the conguration
menu.
Figure 67: Miscellaneous Tab
Touch the AUDIBLE SETTINGS tab.
Press the tab to turn maintenance mode on. Exit the
screens to return to the conguration screen.
Figure 68: Audible Settings Tab
Figure 69: Maintenance Mode ON
Figure 71: Battery Pull Tab

Table of Contents

Other manuals for BeaconMedaes TotalAlert Infinity

Related product manuals