EasyManua.ls Logo

BeaconMedaes TotalAlert Infinity - Gas Combination Alarms; Electromagnetic Compatibility; En 60601-1-2

BeaconMedaes TotalAlert Infinity
122 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1-7
TotalAlert Innity™ HTM Medical Gas Notication System
4107 9016 60.01
1.5.3 Gas Combination Alarms
Figure 1.13: Combination Alarm Front Panel
Gas Combination alarms include a mix of the following
modular components:
B30 board
B60 board
B50 board
Combination alarms can monitor up to 20
BeaconMedaes Line Contact Monitoring signal.
Inputs can be assigned to any one of 8 gas service
indicators.
Combination alarms can also monitor up to 6 digital
gas sensors.
Model Number Scheme:
____ - _______ - ____
Series - Conguration - Retrot
See Model Number Chart in Section 1.7.
Example:
TotalAlert Innity™ Combination Alarm
Example Model Number: A3-C10
1.6 Electromagnetic Compatibility
1.6.1 EN 60601-1-2
Medical Electrical Equipment needs special precautions
regarding EMC and needs to be installed and put into
service according to the EMC information provided in
this manual.
Portable and mobile RF communications equipment
can aect Medical Electrical Equipment. The use of
accessories, transducers, and cables other than those
specied by the manufacturer, may result in increased
emissions or decreased immunity of the
TotalAlert Innity.
The TotalAlert Innity should not be used adjacent
to, or stacked with, other equipment. If adjacent or
stacked use is necessary, the TotalAlert Innity should
be observed to verify normal operation in the
conguration in which it will be used.
Guidance and manufacturer’s declaration -
electromagnetic emissions
The TotalAlert Innity is intended for use in the electromag-
netic environment specied below. The customer or the user
of the TotalAlert Innity should assure that it is used in such an
environment.
Emissions test Compliance
Electromagnetic
environment - guidance
RF emissions
CISPR 11
Group 1
The TotalAlert Innity
uses RF energy only for
its internal function.
Therefore, its RF
emissions are very low
and are not likely to
cause any interference in
nearby electronic
equipment.
RF emissions
CISPR 11
Class A
The TotalAlert Innity
is suitable for use in all
establishments other
than domestic and those
directly connected to the
public low-voltage power
supply network that sup-
plies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage uctuations/
Flicker emissions
IEC 61000-3-3
Complies

Table of Contents

Other manuals for BeaconMedaes TotalAlert Infinity

Related product manuals