4-24
TotalAlert Innity™ HTM Medical Gas Notication System
4107 9016 60.01
Figure 85: Conguration Screen
Figure 84: HTM Signal Conguration Screen
A = Alarm Point
Touch the ALARM POINT tab to select the
proper HTM Signal.
To identify the location where the wires are landed:
1. Touch the tab for the appropriate board and
terminal block. (C1, C2, C3, C4, C5)
2. Touch the Save button when you see a check
beside your selection.
B = Gas Color Code Tab
To select the color code for the appropriate gas based
on your standard and country:
1. Touch the B = GAS COLOR CODE tab (Figure 4.85)
2. Scroll through the gas types using the right and left
arrows at the bottom of the screen.
3. Touch the tab for the appropriate code, or select a
“custom” tab to create a unique source badge.
4. Touch the Save button when you see a check be-
side your selection.
Figure 82: Combination Alarm Screen
Figure 83: Congure Source Screen
4.5.2 Conguring Source Badges
The TotalAlert Innity
™
combination alarm is designed
to allow the conguration of ten source badges. Twenty
locations with up to eight source signals each can be
tied to each source badge.
To congure alarm points in a central badge, touch the
[1] button on the conguration screen (Figure 4.83).
• Congure (Source) Badge