EasyManua.ls Logo

BeaconMedaes TotalAlert Infinity - Customized Instructions Set up

BeaconMedaes TotalAlert Infinity
122 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4-30
TotalAlert Innity™ HTM Medical Gas Notication System
4107 9016 60.01
Audible Settings Tab
The AUDIBLE SETTINGS tab allows for the alarm to
be put in Maintenance Mode. When the Maintenance
mode tab is toggled “ON”, the audible warning will not
reinitialize.
Maintenance mode will automatically change to OFF if
a new alarm condition is introduced or all alarm
conditions are cleared.
Figure 101: Maintenance Mode ON
Figure 102: Maintenance Mode OFF
Figure 103: Active Alarm Opt Tab
Figure 104: General Alarm Message Tab
4.6 Customized Instructions Set Up
Active Alarm Opt Tab
The ACTIVE ALARM OPT tab allows users to customize
their Active Alarm options in two ways: by adding alarm
instructions and by changing the alarm default screen.
Adding Custom Alarm Instructions
Users can add instructions—both general instructions
and specic instructions—to facility personnel for when
the alarm goes into fault status.
To add general instructions:
Touch the GENERAL ALARM MESSAGE tab
Enter the instruction.

Table of Contents

Other manuals for BeaconMedaes TotalAlert Infinity

Related product manuals