4-10
TotalAlert Innity™ HTM Medical Gas Notication System
4107 9016 60.01
Figure 34: Gas Choices
• Touch the right and left arrows at the bottom of
the screen to scroll through additional gas choices,
including options to create custom mixes. Select
the gas that is wired into the D01 input on the GAS
INPUT / B60 board.
• Touch the Save button when you see a check
beside your selection.
• (If an alarm sounds, press the Mute button on the
panel under the screen and continue with setup.)
• Repeat to add additional gases.
• Touch the X button in the upper right to close the
GAS SENSORS screen and return to the PRIMARY
SETTINGS screen. The alarm will retain your infor-
mation.
• Touch the X button in the upper right to close the
PRIMARY SETTINGS screen. The alarm will retain
your information.